The information below will be of help in selecting the correct oils to characterise a viscometer or to calibrate the following flowcups:
| AFNOR | FORD | BS | ISO | DIN | SHELL | ZAHN |
Our Notes on Units may also be useful.
The selection of oils will be governed by the range of viscosities that the equipment is capable of measuring.Our oils can be used for calibration at any temperature in the 20 to 30°C range.
The oils listed in the table below are those that are required to calibrate the specified viscometer throughout its working range. If the viscometer is only used for measurements over a limited range then the oils required for calibration can also be restricted accordingly.
| Viscometer Type | Oil Numbers | |
|---|---|---|
| Brookfield (all models) | Select to cover working range | |
| Cone + plate | 0 – 10 Poise | U 0600 U 1000 U 1400 U 1800 U 2000 |
| U 2200 U 2400 | ||
| 0 – 20 Poise | U 0600 U 1000 U 1400 U 2000 U 2200 | |
| U 2400 U 2600 | ||
| 0 – 40 Poise | U 1000 U 2000 U 22100 U 2400 U 2600 | |
| U 2800 U 3000 | ||
| Krebs Stormer | U 1200 U 1400 U 2000 U 2200 U 2400 | |
| U 2600 U 2800 U 3100 | ||
The following tables list the majority of flow cups in current use together with the oil recommended for characterising them.
These recommendations are based on BS EN ISO 2431:1996, ASTM D 1200 (Re-approved 1999), ASTM 4212-99, BS 3900 A6: 1971 (obsolete), and DIN 53-211: 1987 (obsolete). Obsolete standards are referenced here because they describe cups which are still currently used and are not described in current standards.
Not all flow cups are constructed according to National / International Standards but where possible the table lists the Standard, the equation relating viscosity to flow time and the percentage deviation allowed by the Standard between the measured flow time/viscosity and that calculated using the equation.
| Cup No. | Oil No. | Cup Equation [1]
V = kinematic viscosity (cSt) t = flow time (s) |
Typical flow time (seconds) |
Allowed Tolerance | |
|---|---|---|---|---|---|
| Cup type: AFNOR
Standard: |
4 | – | No data from standard | – | – |
| 6 | – | No data from standard | – | – | |
| Cup No. | Oil No. | Cup Equation [1]
V = kinematic viscosity (cSt) t = flow time (s) |
Typical flow time [2]
(seconds) |
Allowed Tolerance [3] | |
| Cup type: FORD
Standard: |
1 | U 0100 | 33–40 | ±10% | |
| 2 | U 0200 | 61–61 | ±10% | ||
| 3 | U 0200 | 33–36 | ±10% | ||
| 4 | U 0600 | 32–35 | ±10% | ||
| Cup No. | Oil No. | Cup Equation [1]
V = kinematic viscosity (cSt) t = flow time (s) |
Typical flow time [2] (seconds) |
Allowed Tolerance [4] | |
| Cup type: BS
Standard: |
B2 | U 0200 | 146–162 | ±2.5% | |
| B3 | U 0200 | 54–59 | ±2.5% | ||
| B4 | U 0600 | 41–45 | ±2.5% | ||
| B5 | U 1000 | 42–46 | ±2.5% | ||
| B6 | U 2400 | 49–54 | ±2.5% | ||
| Cup No. | Oil No. | Cup Equation [1]
V = kinematic viscosity (cSt) t = flow time (s) |
Typical flow time [2] (seconds) |
Allowed Tolerance [4] | |
| Cup type: ISO
Standard: |
3 mm | U 0100 | 67–70 | ±3% | |
| 4 mm | U 0200 | 48–53 | ±3% | ||
| 5 mm | U 0600 | 35–38 | ±3% | ||
| 6 mm | U 1600 | 54–60 | ±3% | ||
| Cup No. | Oil No. | Cup Equation [1]
V = kinematic viscosity (cSt) t = flow time (s) |
Typical flow time [2] (seconds) |
Allowed Tolerance [4] | |
| Cup type: DIN
Standard: |
4 | U 0600 | 27–29 | ±3% | |
| 6 | – | No data from standard | – | – | |
| 8 | – | No data from standard | – | – | |
| Cup No. | Oil No. | Cup Equation [1]
V = kinematic viscosity (cSt) t = flow time (s) |
Typical flow time [2] (seconds) |
Allowed Tolerance [4] | |
| Cup type: SHELL
Standard: |
1 | – | – | – | |
| 2 | U 0100 | 49–54 | ±20% | ||
| 2.5 | U 0100 | 30–33 | ±20% | ||
| 3 | U 0600 | 74–81 | ±20% | ||
| 3.5 | U 0600 | 51–57 | ±20% | ||
| 4 | U 0600 | 32–36 | ±20% | ||
| 5 | U 0600 | 17–19 | ±20% | ||
| 6 | U 1800 | 27–30 | ±20% | ||
| Cup No. | Oil No. | Cup Equation [1,5]
V = kinematic viscosity (cSt) t = flow time (s) |
Typical flow time [2] (seconds) |
Allowed Tolerance [4] | |
| Cup type: ZAHN
Standard: |
1 | U 0100 | 52–54 | ±20% | |
| 2 | U 0600 | 45–48 | ±20% | ||
| 3 | U 1800 | 44–48 | ±20% | ||
| 4 | U 1800 | 34–37 | ±20% | ||
| 5 | U 2600 | 70–78 | ±20% |